|
During incoming inspection, the distributor rejected this device, 138101, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
|
|
Received four 138101 in unopened original packaging.Lot number was verified.Performed a visual inspection, one end of the packages are not sealed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 68 complaints, regarding 7450 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.003.Per the instructions for use, the user is advised that sterility guaranteed unless package has been opened, broken, or damaged.This issue will continue to be monitored through the complaint system to assure patient safety.
|