| Model Number |
M0062181000 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a passport balloon dilatation catheter was used in the kidney during a ureteroscopy (urs) procedure performed on (b)(6) 2022.During the procedure, physician tried to place the balloon on the guidewire; however, the tip of the balloon got bent.The procedure was completed with another passport balloon dilatation catheter.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: medical device code a040609 captures the reportable event of balloon tip bent.Block h10: investigation result: the returned passport balloon dilatation catheter device was analyzed, and a visual evaluation noted that the balloon was not folded which indicates that the device was subjected to positive pressure.No problems were identified to the balloon material.A visual and tactile evaluation found that the catheter shaft was kinked in multiple sections, indicating that excessive force being applied to the device.It was also observed that the tip of the device was damaged in several location, thereby confirming the reported event.The reported event was confirmed.A labeling review was performed and from the information available, this device was not used per the instructions for use (ifu)/product label.It was reported that the user attempted to load the device into the guidewire.The passport balloon device is not compatible and cannot be loaded into any guidewire.Additionally, the found damage to the shaft and tip is likely due to the excessive force being applied to the device in an attempt to load the device into the guidewire.Taking all available information into consideration, the most probable root cause is failure to follow instructions, since it was determined that the problems traced to the user not following the manufacturer's instructions.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a passport balloon dilatation catheter was used in the kidney during a ureteroscopy (urs) procedure performed on (b)(6) 2022.During the procedure, physician tried to place the balloon on the guidewire; however, the tip of the balloon got bent.The procedure was completed with another passport balloon dilatation catheter.There were no patient complications reported as a result of this event.
|
| |
|