MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7214
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 3.50mm x 20mm nc emerge balloon catheter was selected for use.However, during insertion, the balloon catheter shaft split.The device was pulled out and the procedure was completed with another nc emerge balloon catheter.There were no patient complications nor injuries reported.
 
Event or Problem Description
It was reported that shaft break occurred.A 3.50mm x 20mm nc emerge balloon catheter was selected for use.However, during insertion, the balloon catheter shaft split.The device was pulled out and the procedure was completed with another nc emerge balloon catheter.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.A non-bsc (6f trapliner) guide catheter device was returned separated, and the separated portion of the shaft was still on the nc emerge device.The device was visually and microscopically examined.There were numerous severe hypotube bends and kinks.There were also numerous kinks to the shaft of the device.There was contrast in the inflation lumen and balloon.There was blood in the guidewire lumen and the balloon was loosely folded.The tip of the device was damaged.Product analysis could not confirm reported event as there was no break in the nc emerge balloon catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13548747
Report Number2134265-2022-01719
Device Sequence Number5199400
Product Code LOX
UDI-Device Identifier08714729846673
UDI-Public08714729846673
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2023
Device Model Number7214
Device Catalogue Number7214
Device Lot Number0028484217
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-