| Model Number |
SCF-GWR320-CoCr |
| Medical Device Problem Codes |
Material Fragmentation (1261); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Part scf-gwr320-cocr are k-wires made of cobalt-chrome alloy as per (b)(4): standard specification for wrought cobalt-28chromium-6molybdenum alloys for surgical implants.Considering that this is an implantable material, the risk for the patient is deemed low.
|
| |
|
Event or Problem Description
|
|
During a bilateral insertion of the sliding scfe screw, it was noted after removing the 2.0mm guide wire that the tip of the threaded portion was left in the patient during the second hip.It was not deemed to be in the way of the screw, and the surgeon has decided not to remove it.Both screws were inserted, and the outcome of the surgery was good and as planned.The tip of the wire is outside of the screw, as the distributor understood.The hospital has raised this as a foreign body left-in-patient incident.The wire that broke had been used in the first hip and was being used because the other wires (4 x supplied 2.0mm) were already bent during the surgery.So, it had already been used to insert the first screw.
|
| |
|