(b)(4).Batch #: v96211.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslg2c45a device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All three tones were heard during functional testing (tone 1 is heard when the energy activation button is pressed; tone 2 is heard when tissue impedance threshold is reached and tone 3 is heard when the cycle is complete).No issues were noted with the opening and closing of the jaws.It is possible that excessive body fluids influence the opening and closing of the jaws.The accumulation of body fluids is a routine occurrence during surgical procedures and does not necessarily indicate a manufacturing defect in the device.There were no anomalies noted with the functionality of the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|