| Model Number |
108-SS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient underwent plif spinal fusion surgery (l5-s1) on (b)(6) 2021 to correct grade 3 spondylolisthesis and pain.Two weeks post operatively it was discovered that the left l5 lock screw had loosened.Revision surgery to replace the screw with a larger screw occurred on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Radiographs received confirmed the event.Returned devices were dimensionally measured and found to be in specification.Torque handle was tested and was in specification.Subject device was visually evaluated and marring of the bottom of the set screw that was in contact with the rod, suggests loosening and instability.Final tightening of the lock screw cannot be confirmed from the markings on the bottom of the set screw.Unknown factors include: patient activity at the time or prior to the event, patient bone quality, the degree of spinal instability or patient compliance with post-operative care instructions or if the patient sustained a fall/impact of any sort (no trauma was reported).Root cause is unknown.
|
| |
|