• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PT; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PT; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 2777
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the distal tip of the guidewire detached.A 185cm pt2 moderate support guidewire was selected for procedure.During withdrawal, the distal tip of the guidewire detached.Approximately 1.5 cm distal part of the guidewire remained in the patients coronary artery.The wire has not been in a jail position and kinking has been seen on the coronary artery.The procedure completed with this procedure.There were no further patient complications, and the patient was stable.
 
Event or Problem Description
It was reported that the distal tip of the guidewire detached.A 185cm pt2 moderate support guidewire was selected for procedure.During withdrawal, the distal tip of the guidewire detached.Approximately 1.5 cm distal part of the guidewire remained in the patient's coronary artery.The wire has not been in a jail position and kinking has been seen on the coronary artery.The procedure completed with this procedure.There were no further patient complications, and the patient was stable.It was further reported that the 99% stenosed target lesion in the none tortuous and none calcified proximal intermediate branch.The target lesion itself was not tortuous, but the distal part of the vessel was tortuous.The guidewire was in the distal part of the vessel.The guidewire broken due to the distal part of the vessel was kinking force.The physician did not make any attempts to retrieve the detached guidewire fragment because it was impossible to snare it without the chance of serious complication.The target lesion was treated successfully with another device.The patient is doing well.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: a visual examination of the returned device revealed that the guidewire was found kinked.Microscopic inspection of the guidewire showed the guidewire detached at the distal tip.Dimensional inspection was performed and the distal, medium and proximal of the guidewire were within specifications.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Event or Problem Description
It was reported that the distal tip of the guidewire detached.A 185cm pt2 moderate support guidewire was selected for procedure.During withdrawal, the distal tip of the guidewire detached.Approximately 1.5 cm distal part of the guidewire remained in the patient's coronary artery.The wire has not been in a jail position and kinking has been seen on the coronary artery.The procedure completed with this procedure.There were no further patient complications, and the patient was stable.It was further reported that the 99% stenosed target lesion in the none tortuous and none calcified proximal intermediate branch.The target lesion itself was not tortuous, but the distal part of the vessel was tortuous.The guidewire was in the distal part of the vessel.The guidewire broken due to the distal part of the vessel was kinking force.The physician did not make any attempts to retrieve the detached guidewire fragment because it was impossible to snare it without the chance of serious complication.The target lesion was treated successfully with another device.The patient is doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PT
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13549508
Report Number2134265-2022-01786
Device Sequence Number5023809
Product Code DQX
UDI-Device Identifier08714729471219
UDI-Public08714729471219
Combination Product (Y/N)N
Initial Reporter CountryHU
PMA/510(K) Number
K030617
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2023
Device Model Number2777
Device Catalogue Number2777
Device Lot Number0028258568
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer03/03/2022
03/30/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/21/2022
04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexFemale
-
-