• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number XIFNT610
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant was removed due to an infection at the site.Patient will return for follow up appointment.Tooth #19.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key13549608
Report Number0001038806-2022-00229
Device Sequence Number14749099
Product Code DZE
UDI-Device Identifier00844868022053
UDI-Public(01)00844868022053(17)251030(10)2020100380
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/30/2025
Device Model NumberXIFNT610
Device Catalogue NumberXIFNT610
Device Lot Number2020100380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Initial Report FDA Received Date02/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-