MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS COMMANDER DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS COMMANDER DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9610TF26U
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation of Vessels (2135)
Event Date 01/27/2022
Event Type  Injury  
Manufacturer Narrative
Per the instructions for use (ifu), cardiovascular injuries such as perforation or dissection of vessels, ventricle, myocardial or valvular structures, and hematoma are known potential adverse events associated with the overall thv procedure and may require intervention.According to the thv training manuals, risk factors for aortic dissection, hematoma, or annular rupture during the thv procedure include significant thv over sizing, severely obliterated sinuses of valsalva, porcelain aorta and/or presence of bulky calcification, and narrow calcified stj.In addition, advanced age, female gender, small body weight, and steroid dependency can also be contributing factors.The sapien valve relies on native valve calcium to securely anchor to the annulus.Despite this beneficial aspect of calcium, bulky calcium can increase the risk of calcific nodule displacement into the vasculature, which can lead to vascular injury.At times the extent and distribution of calcium can impair ease of delivery of the valve, correct positioning of the valve, deployment of the valve, and procedural success.Per the instructions for use (ifu), coronary artery perforation is a potential adverse event associated with the transcatheter valve replacement (tvr) procedure and the use of the edwards thv devices.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate (severe calcification and procedure related factors) caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Event Description
Edwards received notification from our affiliate in (b)(6).As reported, this was a case of a 26mm sapien 3 ultra valve implanted in the aortic position via the transfemoral approach.The patient had a very low ejection fraction (25%).The valve was successfully implanted.The degree of valve calcification was severe.After the procedure, the patient's condition got worse.A new angiogram was performed and showed weak coronary bleeding in the left coronary artery.The rupture caused a cardiogenic shock.Therefore, a pci with stent replacement was done.The patient recovered.As per medical opinion, the left coronary artery should have been protected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS COMMANDER DELIVERY SYSTEM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key13550092
MDR Text Key288828062
Report Number2015691-2022-03997
Device Sequence Number1
Product Code NPU
UDI-Device Identifier00690103208078
UDI-Public(01)00690103208078(17)230525(10)63804339
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/25/2023
Device Model Number9610TF26U
Device Catalogue NumberN/A
Device Lot Number63804339
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-