• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ NEEDLE1-ML SYRINGE WITH 25-G X 5/8-IN; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ NEEDLE1-ML SYRINGE WITH 25-G X 5/8-IN; PISTON SYRINGE Back to Search Results
Model Number 305903
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown; awareness date has been used for this field.A sample is available for evaluation.Upon completion of the investigation, a supplemental report will be filed.There were multiple 510k numbers reported to be involved.The information for the additional 510k is as follows: pma / 510(k)#:k951254.
 
Event or Problem Description
It was reported while using bd safetyglide¿ 1-ml syringe with 25-g x 5/8-in needle a stopper was found to be deformed.There was no report of patient impact.The following information was provided by the initial reporter: stopper deformation.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2022-mar-07.Investigation summary: one physical sample and one photo of a 1ml slip tip syringe with safetyglide 25x5/8 needle (p/n 305903) were received.Through visual inspection, it was observed that the stopper was missing from the plunger tip and instead jammed against the inside wall of the barrel and the plunger rod.The observed condition is non-conforming per product specification.Potential root cause for the jammed stopper defect is associated with the assembly process.These conditions are occurring at/below their expected frequency.Therefore, no corrective action is required at this time.Batch #0325431 is considered in compliance with our product specification requirements.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.H3 other text : see h10.
 
Event or Problem Description
It was reported while using bd safetyglide¿ 1-ml syringe with 25-g x 5/8-in needle a stopper was found to be deformed.There was no report of patient impact.The following information was provided by the initial reporter: stopper deformation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAFETYGLIDE¿ NEEDLE1-ML SYRINGE WITH 25-G X 5/8-IN
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13550198
Report Number1213809-2022-00082
Device Sequence Number17439976
Product Code FMF
UDI-Device Identifier30382903059035
UDI-Public30382903059035
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K980580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2025
Device Model Number305903
Device Catalogue Number305903
Device Lot Number0325431
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-