|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Please see the updates in sections d9, h3, h4, h6, and h10.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the probe was found to be broken at the distal end.There was a contact mark on the non-insulated area, wear on the tissue pad, and coating of the grasping section was partially peeled off; however, these defects are not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.Seal & cut¿ output was activated while grasping a thick tissue with the tip of the grasping section.The proximal side of the tissue pad come in contact to the probe.As a result, the pad was worn away.2.The tissue pad was worn away, causing the non-insulated of the grasping section to touch the probe.3.¿seal & cut¿ output was activated under this situation, the scratches indicating that the probe and grasping section were in contact with each other were made.4.The output was continuously activated in state of above description 3, and a force was applied to the probe, causing the probe cracked and the error occurred.5.A load was applied to the probe and it broke.Or 1.The worn of the tissue pad could have happened because ¿seal & cut¿ output was activated while grasping nothing with the grasping section and the probe tip, or kept activating the output after a transection of tissue.2.The tissue pad was worn away, causing the non-insulated of the grasping section to touch the probe.3.¿seal & cut¿ output was activated under this situation, the scratches indicating that the probe and grasping section were in contact with each other were made.This following information is included in the instructions for use (ifu): ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the ptfe pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting or vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the ptfe pad, or the probe tip may break and fall off, and partial separating of the ptfe pad may occur due to a local increase of temperature caused by the friction between ptfe pad and the probe tip during activation.¿ ¿do not activate output while applying the probe tip to the tissue with a strong force, grasping thick tissue, or positioning the tissue, or twisting the shaft, or rotating the rotation knob.Manipulate and isolate the targeted tissue with another instrument, grasp the targeted tissue with the thunderbeat instrument, and then activate.Otherwise, it may cause the deforming/splitting/protruding of tissue pad or a scratch on the probe tip by interference with the other parts, which could result in the probe tip breaking and falling off inside the body cavity.¿ ¿if the grasping section, metal-exposed area around it or the probe tip gets sticked tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ olympus will continue to monitor the performance of this device.
|