Model Number 328431
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 4 bd insulin syringes with bd ultra-fine¿ needles experienced deformed stoppers.The following information was provided by the initial reporter: pet owner reported damaged stopper, 4 syringes were affected.
 
Additional Manufacturer Narrative
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 07-mar-2022.H6: investigation summary: customer returned (4) 0.3ml, 30 gauge, 12.7mm syringes from lot 1172426.Visual inspection found that the stoppers were warped and dragging inside the syringes.Moving the plungers was found to be difficult, with the plungers experiencing more resistance than anticipated.A review of the device history record was completed for batch # 1172426 all inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Based on the samples received, bd was able to confirm the customer¿s indicated failure of stopper damage.Root cause: maintenance dispatch (l2l) #128153 was created for potential dry barrels.Dry barrels indicate a lack of silicone inside the barrel to allow movement of the plunger/stopper rod.Found the rail was not aligned correctly.No root cause as to why the rail was sightly out of line to the silicone being applied into the barrel.H3 other text : see h10.
 
Event or Problem Description
It was reported that 4 bd insulin syringes with bd ultra-fine¿ needles experienced deformed stoppers.The following information was provided by the initial reporter: pet owner reported damaged stopper, 4 syringes were affected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13551185
Report Number1920898-2022-00111
Device Sequence Number5023883
Product Code FMF
UDI-Device Identifier00382908431033
UDI-Public00382908431033
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K024112
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/31/2026
Device Model Number328431
Device Catalogue Number328431
Device Lot Number1172426
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-