MAUDE Adverse Event Report: CORDIS CORPORATION POWERFLEXPRO 8MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION POWERFLEXPRO 8MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400804S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, the balloon of a powerflex pro 8mm4cm 80 percutaneous transluminal angioplasty (pta) dilation catheter ruptured at four atm in an endovascular aneurysm repair procedure.The device was replaced with an unknown balloon catheter to complete the procedure.No injury was reported to the patient.The device was prepped per the instructions for use (ifu).The product was stored properly according to the ifu.There was no difficulty removing the product from the hoop or removing the protective balloon cover.There was no difficulty removing the stylet or any of the sterile packaging components.The device prepped normally and maintained a negative pressure.There were no kinks or other damages noted prior to inserting the product into the patient.Other procedural details were requested but not provided.The product was not returned for analysis as it was discarded due to suspicion of infectious disease.A product history record (phr) review of lot 82178227 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Event or Problem Description
As reported, the balloon of a powerflexpro 8mm4cm 80 percutaneous transluminal angioplasty (pta) dilation catheter ruptured at 4atm in an endovascular aneurysm repair procedure.The device was replaced with an unknown balloon catheter to complete the procedure.No injury was reported to the patient.The device was prepped per the instructions for use (ifu).The product was stored properly according to the ifu.There was no difficulty removing the product from the hoop, or removing the protective balloon cover.There was no difficulty removing the stylet or any of the sterile packaging components.The device prepped normally and maintained a negative pressure.There were no kinks or other damages noted prior to inserting the product into the patient.Other procedural details were requested but not provided.The device will not be returned for evaluation as it was discarded due to suspicion of infectious disease.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 8MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key13551596
Report Number9616099-2022-05356
Device Sequence Number17399346
Product Code LIT
UDI-Device Identifier20705032059814
UDI-Public(01)20705032059814(17)220831(10)82178227
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number4400804S
Device Catalogue Number4400804S
Device Lot Number82178227
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/21/2022
Initial Report FDA Received Date02/17/2022
Date Device Manufactured09/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN BALLOON CATHETER
Patient SexUnknown
-
-