MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 3542625
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Crushing Injury (1797)
Date of Event 10/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a (b)(6) caucasian female patient underwent breast augmentation revision with two 200cc mentor siltex round moderate profile saline breast prostheses.Post-operatively, the patient was physically diagnosed with right breast implant deflation.It was indicated that the deflation was right after a mammogram.As a result, the patient underwent implant removal on (b)(6) 2022.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: material rupture after mammogram.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: material rupture after mammogram.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) caucasian female patient underwent breast augmentation revision with two 200cc mentor siltex round moderate profile saline breast prostheses.Post-operatively, the patient was physically diagnosed with right breast implant deflation.It was indicated that the deflation was right after a mammogram.As a result, the patient underwent implant removal on (b)(6) 2022.
 
Additional Manufacturer Narrative
On march 21, 2022, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the sal siltex rnd diap pkg 200cc breast implant had a tear on the posterior view, measuring approximately 0.5 cm.The evaluation determined that the cause of the deflation is the trauma experienced.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of the saline-filled mammary prosthesis.A second product was received (5543685).No adverse events were reported for this concomitant (contralateral) device, therefore no further analysis is required.It should be noted that as part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.
 
Additional Manufacturer Narrative
On february 28, 2022, mentor became aware that the patient underwent bilateral removal and replacement with the following devices: (right) 125cc mentor smooth round moderate profile saline catalog: 3501610 lot: 9665537 sn: (b)(6) and (left) 125cc mentor smooth round moderate profile saline catalog: 3501610 lot: 9665537 sn: (b)(6) a manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.On march 10, 2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13552307
Report Number1645337-2022-01765
Device Sequence Number5199493
Product Code FWM
UDI-Device Identifier00081317001744
UDI-Public00081317001744
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2008
Device Model Number3542625
Device Catalogue Number3542625
Device Lot Number5528390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer02/28/2022
03/21/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/20/2022
03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2004
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-