Model Number HC-1500
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Abrasion (1689)
Date of Event 01/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the unit and found that the screws that attach the panel to the sterilizer were loose causing the reported event to occur.The technician installed new screws, tested the unit, confirmed it to be operating according to specification and returned it to service.No additional issues have been reported.
 
Event or Problem Description
The user facility reported that the top panel on their evolution sterilizer fell off of the unit contacting an employee's head.Medical treatment was administered (skin adhesive).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
66" EVOLUTION STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key13552322
Report Number3005899764-2022-00005
Device Sequence Number12062083
Product Code FLE
UDI-Device Identifier00724995151522
UDI-Public00724995151522
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHC-1500
Device Catalogue NumberUE06
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2022
Initial Report FDA Received Date02/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-