• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEGREE XLPE LNR 32MM X 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEGREE XLPE LNR 32MM X 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71339554
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a thr on (b)(6) 2021, the patient experienced recurrent dislocation in the joint.A revision surgery was performed on (b)(6) 2022 to revised an r3 0 degree xlpe lnr 32mm x 54mm and oxinium fem hd 12/14 32mm +8 to correct this issue.The current health status of patient is unknown.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a thr on (b)(6) 2021, the patient experienced recurrent dislocation in the joint.A revision surgery was performed on (b)(6) 2022 to revised an r3 0 degree xlpe lnr 32mm x 54mm and oxinium fem hd 12/14 32mm +8 to correct this issue.The current health status of patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated dislocation.The clinical/medical investigation concluded that, two, undated x-rays provided were reviewed.However, the x-rays alone do not contribute to the clinical/medical investigation of the root cause of the recurrent dislocations.No other medical/ clinical documentation was provided.Without supporting medical documentation, a thorough medical assessment cannot be performed.In the event additional medical/clinical records are received, the clinical task may be re-opened, and a thorough assessment will be rendered at that time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device revealed that dislocation has been identified in adverse events in primary and revision surgery.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEGREE XLPE LNR 32MM X 54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13552331
Report Number1020279-2022-00770
Device Sequence Number5199931
Product Code JDI
UDI-Device Identifier03596010583093
UDI-Public03596010583093
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71339554
Device Catalogue Number71339554
Device Lot Number19KM15290
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer03/18/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/22/2022
Date Device Manufactured10/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-