EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PXVJ0711
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Udi number is (b)(4).Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event or Problem Description
It was reported that a vamp jr.Tubing was broken and detached near the distal end of the vamp draw port upon removal from the package.Nurse noticed line set-up was completely separated into 2 pieces.Separation was distal to the vamp draw port just prior to the proximal 3-way stopcock.No patient injury reported.Patient information was not shared.
 
Additional Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Udi number is (b)(4).Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event or Problem Description
It was reported that a vamp jr.Tubing was broken and detached near the distal end of the vamp draw port upon removal from the package.Nurse noticed line set-up was completely separated into 2 pieces.Separation was distal to the vamp draw port just prior to the proximal 3-way stopcock.No patient injury reported.Patient information was not shared.
 
Additional Manufacturer Narrative
The report of tubing breakage was confirmed.Pressure tubing was found completely broken from the bond joint with female connector at distal end of vamp jr system.Cross surfaces of broken tubing appeared rough and uneven.No other visible damage was observed from returned kit.Per product evaluation findings the tube was broken and not detached, therefore this is no longer a reportable event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Common Device Name
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key13552356
Report Number2015691-2022-04009
Device Sequence Number14701946
Product Code DXO
UDI-Device Identifier57460691954484
UDI-Public(01)57460691954484(17)230125(11)210903(10)63937398
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K925638
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date01/25/2023
Device Model NumberPXVJ0711
Device Lot Number63937398
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-