| Model Number |
24706 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the anterior tibial artery.A 2.5mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during inflation at 14 atmospheres, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: updated event description.Device evaluated by mfr.: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.The balloon was inflated to rated burst pressure and was able to hold pressure.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis could not confirm the reported rupture.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the anterior tibial artery.A 2.5mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during inflation at 14 atmospheres, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient was stable.It was further reported that the target lesion was located in the severely tortuous and severely calcified anterior tibial artery.The balloon was inflated 2-3 times at 14 atmospheres and was completely removed.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the anterior tibial artery.A 2.5mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during inflation at 14 atmospheres, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.The balloon was inflated to rated burst pressure and was able to hold pressure.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis could not confirm the reported rupture.
|
| |
|