| Model Number |
1806-2035 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported a missed target was discovered when checking the target accuracy before nail insertion.Spare instruments were arranged and used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since physical evaluation revealed a targeting inaccuracy.The device inspection revealed the following: the received targeter shows traces of a frequent and intense use.The printings on the target device were turned from white to fawn; an indication for a high level of sterilization cycles.Detailed inspection revealed that the distal shaft is slightly twisted in itself.Simulation of a pre-functional check revealed that the function of the item is no longer given due to damage verified.The reported targeting inaccuracy was reproducible with sample counterparts.Based on investigation and since nothing was reported after the last usage in a surgical procedure and furthermore, during assembling before sterilization in the current case, the root cause is rather related to an inadequate reprocessing.Known from similar cases the bent shaft was caused by a result of bending forces during reprocessing, most likely caused by loaded instruments on the targeter.Nevertheless, as instructed per ifu accuracy in guidance needs to be checked prior to the surgical procedure.If simulation of a pre-operative check would have been performed before the surgical procedure and after reprocessing performed it would have been noticed at this time as it¿s clear evident.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
|
| |
|
Event or Problem Description
|
|
It was reported that a missed target was discovered when checking the target accuracy before nail insertion.Spare instruments were arranged and used to complete the procedure but the surgery was interrupted for almost two and a half hours and patient required additional anesthesia.
|
| |
|