| Model Number |
TMICL12.6 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight: unk.Ethnicity: unk.Race: unk.Work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.6mm tmicl12.6 implantable collamer lens, -10.00/+3.5/091 (sphere/cylinder/axis), in the patients left eye (os) on (b)(6) 2021.The lens rotated at post-op day 1 and was explanted on (b)(6) 2022.The lens was replaced with a longer lens but the problem was not resolved, the lens rotated again.See mfr.Report # 2023826-2022-00484 for replacement lens.
|
| |
|
Additional Manufacturer Narrative
|
|
H3- device evaluation: the lens was returned in a specimen cup in liquid.Visual inspection found no visble damage to the lens.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|