| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -09.50/+3.5/085 (sphere/cylinder/axis), in the patients left eye (os) on (b)(6) 2022.The lens rotated at post-op day 1 and was explanted on (b)(6) 2022.The lens was replaced with a non toric lens.See mfr.Report # 2023826-2022-00483 for the initial lens.The patient will eventually require prk surgery for his astigmatism because his eye anatomy is not conducive to an icl remaining on axis.
|
| |
|
Additional Manufacturer Narrative
|
|
Weight: unk.Ethnicity: unk.Race: unk.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3- device evaluation: the lens was returned in a specimen cup in liquid.Visual inspection found no damage to the lens.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|