| Model Number |
3425-0340S |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient fell into the gutter after surgery.It was noted that the nail may have been fractured.Pseudoarthrosis was also identified.The patient underwent the revision surgery on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient fell into the gutter after surgery.It was noted that the nail may have been fractured.Pseudoarthrosis was also identified.The patient underwent the revision surgery on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the received information, the root cause was attributed to a patient related issue.The information indicates that a combination between a implantation time of more than one year, the reported pseudoarthrosis and finally the fall did lead to the breakage of the device.In this relation following statement of the instructions for use can be mentioned: [adverse effects conditions attributable to non-union, osteoporosis, osteomalicia, diabetes, inhibited revascularization and poor bone formation can cause loosening, bending, cracking, fracture of the device or premature loss of rigid fixation with the bone.] [original statements] if more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient fell into the gutter after surgery.It was noted that the nail may have been fractured.Pseudoarthrosis was also identified.The patient underwent the revision surgery on (b)(6) 2022.
|
| |
|