Model Number 3425-0340S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 01/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the patient fell into the gutter after surgery.It was noted that the nail may have been fractured.Pseudoarthrosis was also identified.The patient underwent the revision surgery on (b)(6) 2022.
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the patient fell into the gutter after surgery.It was noted that the nail may have been fractured.Pseudoarthrosis was also identified.The patient underwent the revision surgery on (b)(6) 2022.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the received information, the root cause was attributed to a patient related issue.The information indicates that a combination between a implantation time of more than one year, the reported pseudoarthrosis and finally the fall did lead to the breakage of the device.In this relation following statement of the instructions for use can be mentioned: [adverse effects conditions attributable to non-union, osteoporosis, osteomalicia, diabetes, inhibited revascularization and poor bone formation can cause loosening, bending, cracking, fracture of the device or premature loss of rigid fixation with the bone.] [original statements] if more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that the patient fell into the gutter after surgery.It was noted that the nail may have been fractured.Pseudoarthrosis was also identified.The patient underwent the revision surgery on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø10X340MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13552761
Report Number0009610622-2022-00058
Device Sequence Number5186581
Product Code HSB
UDI-Device Identifier04546540375346
UDI-Public04546540375346
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/31/2023
Device Model Number3425-0340S
Device Catalogue Number34250340S
Device Lot NumberK0B7B56
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/21/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-