| Model Number |
545-511 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Great Vessel Perforation (2152); Cardiac Perforation (2513)
|
| Date of Event |
01/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient's date of birth unk.Patient's weight unk.Relevant tests/laboratory data unk.Device lot number, expiration date unk.The device was discarded, thus no investigation could be completed.Device manufacture date unk because lot number unk.
|
| |
|
Event or Problem Description
|
|
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to infection.Spectranetics lead locking devices (llds) were inserted into each lead to provide traction.The physician started removal of the rv lead using a spectranetics tightrail rotating dilator sheath.Progress was made, with the device advancing down to the superior vena cava (svc)/ra junction when the patient's blood pressure suddenly dropped.Rescue efforts began immediately; ct surgeon was already in the room, and sternotomy was performed.A 2 cm perforation in the svc/ra junction was discovered.Repair of the tear was successful.Rv and ra leads were removed post sternotomy and patient survived the procedure.This report captures the tightrail device in use when the svc/ra junction perforation occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|
| |
|