MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA 5.5MM W/2 UB (WHT & BLUE); FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA 5.5MM W/2 UB (WHT & BLUE); FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202895
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an arthroscopy procedure, two twinfix ultra screws were fractured during screwing.The procedure was successfully completed with a non-significant delay using a back-up device.No patient complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy procedure, two twinfix ultra screws were fractured during screwing.All the broken pieces were retrieved from the patient by tweezers.The procedure was successfully completed with a non-significant delay using a back-up device in the originally drilled bone hole.No patient complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A review of the customer provided image found labelling confirming the product identification information.There are two anchors pictured, and one appears to have some debris in the threads and on the proximal end.No physical deficiencies can be seen on either anchor.The complaint was not confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest that the anticipated risk region is not adequate.A review of the anchor found that found that no burrs, cracks, or contamination is allowed.A certification of compliance to material and processing specifications is required.Based on the information provided, all the broken pieces were removed from inside of the patient using tweezers.According to the report, the surgeon completed the procedure with a non-significant delay using a back-up device.Since there was no alleged harm to the patient, no further clinical/medical assessment is warranted at this time.Should any additional relevant clinical information be provided, this case will be re-assessed.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA 5.5MM W/2 UB (WHT & BLUE)
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13553079
Report Number1219602-2022-00247
Device Sequence Number14407550
Product Code MBI
UDI-Device Identifier03596010652263
UDI-Public03596010652263
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K100159
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/07/2025
Device Model Number72202895
Device Catalogue Number72202895
Device Lot Number2048337
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer02/24/2022
04/26/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date02/28/2022
04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight46 KG
-
-