Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2022
Event Type  malfunction  
Event Description
The user facility reported to terumo cardiovascular, prior to cardiopulmonary bypass, during prime, leakage was observed from the shunt sensor.Per facility, the tube was changed but the leakage was still observed.The priming solution that was voluven solyugen medication.No patient involvement.The product was changed out.The surgery was completed successfully.
 
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 18, 2021.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 213, 19).Type of investigation: #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation finding: 213 - no device problem found.Investigation conclusions: 19 - cause traced to user.The returned sample was visually inspected upon receipt, during which it was noted that the sparger assembly and white luer cap were not present.A white luer cap was added to the sample.The returned sample was then leak tested, as received by connecting with a calibrated manometer, submerged into a water bath, and pressurized up to 1000 mmhg and a leak was noted immediately.The large bore adapter blue cap was then loosened and re-tightened by hand.The sample was then leak tested a second time by connecting with the calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg, and no leaks were noted.A retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside the pouch.The retention sample was then pressurized with air up to 1000 mmhg, submerged in a water bath, and observed for any leaks.No leaks were noted on the retention sample.The root cause for this event was determined to be the large blue vent cap for shunt sensor was not fully tightened either during setup of the circuit, or after the gas calibration.When the large blue vent cap was loosened, it had not been re-tightened fully prior to use in the line, causing a leak from the cap.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key13558174
MDR Text Key285809207
Report Number1124841-2022-00026
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753160767
UDI-Public(01)00699753160767
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberZM11F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-