Model Number N/A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Initial reporter occupation: cath lab director.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
 
Event or Problem Description
The user facility reported that a 7 fr precision sheath was used via jugular access for a heart failure case.While the sheath was being removed, the tip broke off.The tip of the sheath was retrieved with a snare.The patient was stable.The procedure was completed.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed since the device was not available for evaluation.The exact root cause cannot be determined.Review of dhr showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the fmea.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE PRECISION ACCESS SYSTEM
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
stephanie handy
950 elkton blvd.
elkton, MD 21921
9499890491
MDR Report Key13558283
Report Number1118880-2022-00007
Device Sequence Number4162324
Product Code DYB
UDI-Device Identifier00389701010496
UDI-Public00389701010496
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K111606
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2023
Device Model NumberN/A
Device Catalogue Number70-7130
Device Lot Number0000022047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer03/04/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SNARE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-