The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.As the device had signs of degradation, the reported event is confirmed to have a relationship with the device.As the cause of this issue is unknown, it cannot be determined if the device met specifications.As the product was not used on a patient, it was not used for diagnostic or treatment purposes.An applicable risk region and failure mode will be addressed in a future revision of afmea.A red ureteral catheter was returned opened in original packaging.The catheter was noted to contain white discoloration and white striped discoloration caused by oxidation of the catheter.As is it unable to determined when the oxidation occurred the reported event will be confirmed cause unknown.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.
|