• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA®; DIMENSION VISTA® MAGNESIUM (MG) FLEX® REAGENT CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA®; DIMENSION VISTA® MAGNESIUM (MG) FLEX® REAGENT CARTRIDGE Back to Search Results
Model Number 10445158
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A united states customer contacted the siemens healthcare diagnostics remote services center (rsc) regarding a discordant falsely depressed patient sample result on a dimension vista system.There were no known reports of adverse health consequences or potential for serious injury to the patient due to the discordant magnesium result or treatment administered.This mdr is being filed conservatively as the patient received magnesium.Siemens healthcare diagnostics headquarters support center (hsc) concluded the investigation of the falsely depressed patient sample magnesium result on a dimension vista system.Siemens healthineers has confirmed that the dimension vista magnesium (mg) flex reagent cartridge lots 21253bb, 21263ba and 21335aa may show higher imprecision across the analytical measurement range for quality control and patient samples.Customers have reported a maximum negative difference of (b)(6) mg/dl (-0.9 mmol/l) between two replicates and maximum positive difference between two replicates to be 1.5 mg/dl (0.6 mmol/l).Urgent medical device correction vc-22-01.A.Us was issued to united states customers and urgent field safety notice (ufsn) vc 22-01.A.Ous was sent to outside the us (ous) customers advising customer to discontinue use and discard these dimension vista magnesium lots.
 
Event or Problem Description
A discordant, falsely depressed magnesium (mg) patient sample result was obtained on a dimension vista 1500 system.The result was reported to the physician.The same sample was reprocessed on an alternative dimension vista system and a higher result, considered correct, was obtained.A corrected report was issued.The patient was administered magnesium based on the discordant result.There were no known reports of adverse health consequences to the patient due to the discordant magnesium result or treatment.
 
Additional Manufacturer Narrative
Additional information (18-feb-2022).Siemens healthcare diagnostics, inc.Filed corrections and removal report (crr) dated 08-feb-2022 with the fda per 21 cfr 806.10.Root cause investigation is in progress and corrective action updates will be provided in monthly status update reports to the fda as per 21 cfr part 806.10.Initial mdr 2517506-2022-00031 was filed 18-feb-2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA®
Common Device Name
DIMENSION VISTA® MAGNESIUM (MG) FLEX® REAGENT CARTRIDGE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer Contact
james morgera
500 gbc drive
po box 6101
newark, DE 19714-6101
MDR Report Key13558843
Report Number2517506-2022-00031
Device Sequence Number11709451
Product Code JGJ
UDI-Device Identifier00842768015694
UDI-Public00842768015694
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140790
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2022
Device Model Number10445158
Device Catalogue NumberK3057 SMN 10445158
Device Lot Number21335AA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2517506-02/08/2022-001-C
Patient Sequence Number1
Patient Age19 YR
Patient SexMale
-
-