Catalog Number 8801-90034
Medical Device Problem Codes Fracture (1260); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of an ozark screw driver wouldn't retain the screw intra-operatively.Surgery was successfully completed with no delay.No adverse consequences or medical intervention were reported.
 
Additional Manufacturer Narrative
A visual inspection was performed.The threaded tip of the driver has fractured off.The threaded portion fracture is just proximal to the tip of the hexalobe.Device history records were reviewed for this lot, and no relevant manufacturing issues were identified.Complaint history records were reviewed for this lot and catalog number, no adverse trends were identified.Initial report states that the driver will not retain the screw and visual inspection confirmed the threaded tip of the driver has fractured off.It was also reported via follow up that the blue handle was used with the driver during screw placement.The device was not used at a difficult angle or with excessive force.The most likely cause of the reported event was determined to be due to the use of the incorrect handle during screw insertion.The grey torque limiting handle must be used with the threaded driver.The torque limiting handle is designed to protect the driver from excessive torque, which can lead to driver damage during screw insertion.
 
Event or Problem Description
It was reported that the tip of an ozark screw driver wouldn't retain the screw intra-operatively.Surgery was successfully completed with no delay.No adverse consequences or medical intervention were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW DRIVER W/ THREAD; SIZE 10
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13559477
Report Number3004774118-2022-00067
Device Sequence Number3818822
Product Code LXH
UDI-Device Identifier10888857400863
UDI-Public10888857400863
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8801-90034
Device Lot NumberJDCR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1