• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. OPUS SPEEDSCREW IMPLANT; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. OPUS SPEEDSCREW IMPLANT; SCREW, FIXATION, BONE Back to Search Results
Model Number OM-6500
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/06/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an arthroscopy, a little piece of metal of the opus speedscrew anchor broke, and it was left inside of the patient because the removal process would have done more damage than good.It is unknown how the procedure was completed and if there was a delay.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Additional Manufacturer Narrative
H2: additional information ¿b5: event description¿.
 
Event or Problem Description
It was reported that during an arthroscopy, a little piece of the opus speedscrew anchor broke.The piece was left inside of the patient because the removal process would have done more damage than good.The procedure was completed with non-significant delay using a s+n back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H10 h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A review of the material specifications, found that tensile strength should be verified to meet specification of 100 - 118 mpa, and a certificate of material analysis is required with each shipment.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Based on the information provided, ¿the surgeon chose to leave the little piece of the opus speed screw anchor inside of the patient instead of trying to remove it because it would do more harm to the patient trying to remove it.¿ the retained screw is implantable.So biocompatibility is not an issue.Although, we cannot rule out the possibility of micro-motion and/or migration of the retained piece.According to the report, the procedure was completed with a non-significant delay using a s&n back-up device.Since no further complications were reported.No further clinical medical assessment is warranted at this time.Should any additional relevant patient information be provided, this case would be re-assessed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPUS SPEEDSCREW IMPLANT
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13559488
Report Number3006524618-2022-00061
Device Sequence Number19766616
Product Code HWC
UDI-Device Identifier00817470005196
UDI-Public00817470005196
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K081893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOM-6500
Device Catalogue NumberOM-6500
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer02/23/2022
11/17/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date02/28/2022
11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age65 YR
Patient SexMale
Patient Weight95 KG
-
-