BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 60% stenosed target lesion was located in a non-tortuous and non-calcified fistula of the left arm.A 8x40x75 epic vascular stent was advanced and deployed to treat the lesion.However, during removal, it was noticed that the inner white shaft was separated and was moving freely from the handle.The shaft did not dislodge entirely from the stent delivery system.The device was removed with ease and the procedure was completed.There were no patient complications.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the inner liner is separated from the handle.The sheath is kinked 34cm and bent 3.5cm from the strain relief.There are multiple kinks to the inner liner.Microscopic examination revealed no additional damages.Because there was no evidence of any product quality deficiencies, it was considered likely that the kinks were attributable to handling.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that shaft break occurred.The 60% stenosed target lesion was located in a non-tortuous and non-calcified fistula of the left arm.A 8x40x75 epic vascular stent was advanced and deployed to treat the lesion.However, during removal, it was noticed that the inner white shaft was separated and was moving freely from the handle.The shaft did not dislodge entirely from the stent delivery system.The device was removed with ease and the procedure was completed.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13559614
Report Number2134265-2022-01126
Device Sequence Number3818829
Product Code NIO
UDI-Device Identifier08714729805656
UDI-Public08714729805656
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/23/2026
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0027553566
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer02/22/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-