• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL Back to Search Results
Model Number M006180155010
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a contour vl ureteral stent was used during a stent placement procedure in the ureter.The event date was not reported.During preparation, it was noted that the contour vl ureteral stent was found separated.The procedure was successfully completed with another contour vl ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be as no patient injury.
 
Additional Manufacturer Narrative
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Initial reporter state: (b)(6).Medical device code a0401 captures the reportable event of stent shaft broken.The returned contour vl ureteral stent was analyzed, and a visual evaluation noted that the bladder pig tail torn and drag marks noted in the device.No other problems with the device were noted.The reported event of stent shaft break was not confirmed.Based on the most relevant information of the complaint event description, device analysis and including the product record review results, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It was found that the device has the bladder pig tail torn and drag marks noted in the device.According to the evidence, it is possible that operational factors, interaction of the device with the suture string such as entanglement, manipulation or excess of force applied during it use, could have caused the torn and drag marks observed.Consequently, affect the performance of the device.Therefore, the most probable cause of this complaint is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a contour vl ureteral stent was used during a stent placement procedure in the ureter.The event date was not reported.During preparation, it was noted that the contour vl ureteral stent was found separated.The procedure was successfully completed with another contour vl ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be as no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR VL
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13559738
Report Number3005099803-2022-00532
Device Sequence Number3734279
Product Code FAD
UDI-Device Identifier08714729058427
UDI-Public08714729058427
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K974541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2022
Device Model NumberM006180155010
Device Catalogue Number180-155-01
Device Lot Number0023107855
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-