Model Number SA60WF
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
A health care professional reported during the intraocular lens implantation the lens was faulty, melted and deformed as a result it could not be used.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.There are no other complaints for the reported lot.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.There are no other complaints for the reported lot.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional reported during the intraocular lens implantation the lens was faulty, melted and deformed as a result it could not be used.Additional information has been requested.
 
Event or Problem Description
Additional information has been received and stated that, the lens was faulty, melted and deformed was found before the surgery started.There was no patient contact noted.Based on information received following submission of the previous report, this event does not meet criteria for reporting as a malfunction.
 
Additional Manufacturer Narrative
Based on information received following submission of the previous report, this event does not meet criteria for reporting as a malfunction.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported damage was observed.Iol returned pressed down on two(2) posts of the lens case base resulting in gross optic and haptic damage.Solution is dried on both surfaces of the optic and one haptic.One haptic is deformed and has haptic tip damage rejectable, the other haptic is nicked/chipped-rejectable.The optic is deformed, scratched rejectable.We are unable to determine the root cause for the reported complaint - "fault on the lens, lens is melted and deformed".The returned iol shows evidence of possible handling by the customer due to the presence of solution dried on both surfaces of the optic and one haptic.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13560135
Report Number9612169-2022-00076
Device Sequence Number17110786
Product Code HQL
UDI-Device Identifier00380652332682
UDI-Public00380652332682
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/22/2026
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.210
Device Lot Number25273659
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer04/11/2022
06/02/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date05/07/2022
06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-