C.R. BARD, INC. (COVINGTON) -1018233 BARD® RUBBER UTILITY CATHETER Back to Search Results
Model Number 277716
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the patient was having drainage issues with the use of all purpose catheter.Patient stated that the eyelets were too small and not draining all of the urine from bladder.No additional information provided.Per additional information received via liberator on 03feb2022, it was stated that all purpose catheter felt like fiberglass and it causes bleeding.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A potential root cause for this failure could be "operator error, dull punch dies".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "rubber utility catheter; x-ray opaque rubber.Warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original.Caution: this product contains natural rubber latex which may cause allergic reactions.Caution: federal (usa) law restricts this device to sale by or on the order of a physician." correction: d.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the patient was having drainage issues with the use of all purpose catheter.Patient stated that the eyelets were too small and not draining all of the urine from bladder.No additional information provided.Per additional information received via liberator on (b)(6) 2022, it was stated that all purpose catheter felt like fiberglass and it causes bleeding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® RUBBER UTILITY CATHETER
Common Device Name
RUBBER UTILITY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13560190
Report Number1018233-2022-00717
Device Sequence Number4313123
Product Code KOD
UDI-Device Identifier00801741034114
UDI-Public(01)00801741034114
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2026
Device Model Number277716
Device Catalogue Number277716
Device Lot NumberNGFX1355
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date02/18/2022
Supplement Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-