Catalog Number 955467 |
Device Problems
Disconnection (1171); Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/23/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that after one hour of treatment with a prismaflex m150 set c, an external fluid leak was observed due to a separated drain bag stopcock.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Facility name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Facility name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that after one hour of treatment with a prismaflex m150 set c, an external fluid leak was observed due to a separated drain bag stopcock.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the pictures showed that the stopcock of the drain bag was disconnected.The reported condition was verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|