| Model Number |
3010-19-008 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6)2022 during spinal fusion while attempting to navigate the verse pedicle screw, the holding sleeve would not retain on the driver.This would not allow screw to be navigated.Navigation of screw was discontinued and screw inserted over guidewire.Patient status is unknown.There was no surgical delay.The procedure was successfully completed.This report involves one (1) navigation enabled instruments expedium verse driver - shaft.This is report 1 of 2 for (b)(4).
|
| |
|