| Model Number |
04.043.440S |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional product code: hwc.Complainant part is not expected to be returned for manufacturer review/ investigation.Part 04.043.440s, lot 179p269: release to warehouse date: june 25, 2021.Expiration date: may 31, 2031.Manufactured by: (b)(4).A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, the patient with a tibia shaft fracture was being treated with a suprapatellar tibial nail.During insertion of the tibial nail, the protection sleeve inner metal portion became delaminated from the plastic portion of the sleeve.This was identified prior to locking screw insertion.The nail was explanted, the pieces of the sleeve were removed, and the nail was reinserted.There were fifteen minutes surgical delay.The procedure was successfully completed.There was no patient consequence reported.This report is for a tibial nail.This is report 2 of 2 for (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|