The reporter indicated the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, 07.50/+1.5/095 (sphere/cylinder/axis) in the patient's right eye (od) on (b)(6) 2022.The lens was explanted on (b)(6) 2022 due to excessive vaulting, angle closure with elevated intraocular pressure (iop) and significant reduction of irido-corneal angles.There was an addition of new pi by yag.The lens was replaced with a shorter lens and the problem was resolved.Vault and iop at normal post icl target with smaller icl.At one week post-op, there was trace cell and spk.The cause of the event was unknown.
|