| Model Number |
68-750 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed, an opening in the radius side assessed to a fold flaw opening and have four openings in the anterior side due to surgical damage like.Additional observations: crease fold observed in the device.Brown particles inside the device.No further actions are required as the device was implanted.
|
| |
|