• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number PCB00
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the iol (intraocular lens) was implanted, but appeared to flip which caused a capsular rupture and required the vitrectomy probe to be used.Although the capsule was stable, the surgeon decided to remove the iol.It was decided not to implant another iol until the eye had settled.The lens was injected as normal and the surgeon was unsure why it had turned/ flipped.The lens was inserted and removed during same procedure.There was a 10 minute delay in the procedure.There was no incision enlargement or sutures required.The patient's post-operative outcome was not available.No further information is available.
 
Additional Manufacturer Narrative
Age, weight and ethnicity: per eu regulation (b)(4) (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted; give date: n/a (not applicable).The intraocular lens was inserted and removed during the same procedure.If explanted; give date: n/a (not applicable).The intraocular lens was inserted and removed during the same procedure.(b)(6).Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record evaluation: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be confirmed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13564585
Report Number3012236936-2022-00447
Device Sequence Number5848715
Product Code HQL
UDI-Device Identifier05050474558489
UDI-Public(01)05050474558489(17)231130
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberPCB00
Device Catalogue NumberPCB0000310
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Initial Report FDA Received Date02/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-