• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LNR, 10 DEG FACE, 32MM ID, GRP 1 CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LNR, 10 DEG FACE, 32MM ID, GRP 1 CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number NV GXL LNR, 10 DEG FACE, 32MM ID, GRP 1 CUP
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately ten years post initial tha, this female patient's poly broke.Patient was revised to an xle liner.There were no broken pieces fell in the patient wound site.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Photos and x-ray were unable to obtain.Device is not available for evaluation as device was kept by hospital and thrown away.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the most likely underlying cause of the prosthesis fracture and the type of fracture reported cannot be determined from the information provided.It is unclear if the acetabular liner locking button failed, if the liner wore through, or if there was a fracture around the liner rim.The most probable root cause associated with the reported event of "prosthesis fracture¿ is associated with a partial or full-thickness crack in the device materials, either during implantation or sometime after.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, 10 DEG FACE, 32MM ID, GRP 1 CUP
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13565337
Report Number1038671-2022-00205
Device Sequence Number14939167
Product Code JDI
UDI-Device Identifier10885862024732
UDI-Public10885862024732
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/12/2017
Device Model NumberNV GXL LNR, 10 DEG FACE, 32MM ID, GRP 1 CUP
Device Catalogue Number138-32-51
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date02/19/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/14/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-