Model Number 470183-14
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/27/2022
Event Type  Injury  
Manufacturer Narrative
Intuitive surgical, inc.(isi) has not received the permanent cautery hook instrument for evaluation.Therefore, the root cause of the customer-reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.A review of the site's complaint history does not show any additional complaints related to this product.An image of the permanent cautery hook (pch) instrument was received from the customer.A review of the provided image was consistent with the report of a plastic piece falling off the instrument.Further details regarding the failure mode will be obtained through the failure analysis investigation upon the return of the instrument.A review of the system log and instrument log was performed.The instrument was last used on (b)(6) 2022 on system (b)(4).The permanent cautery hook has 2 uses remaining after the last use.The permanent cautery hook has a maximum of 10 uses.This complaint is being reported due to the following conclusion: during a da vinci-assisted surgical procedure, it was alleged that a permanent cautery hook (pch) instrument broke and a plastic piece was missing.The plastic piece fell inside of the patient and the surgeon was unable to find a missing piece of ultem.It is unknown what caused the breakage to occur.Additionally, it is unknown if the fragment was retrieved or if the fragment remains inside the patient's anatomy.
 
Event Description
It was reported that during a da vinci-assisted hysterectomy surgical procedure, a permanent cautery hook (pch) instrument broke and a plastic piece was missing.The plastic piece fell inside of the patient and the surgeon was unable to find a missing piece of ultem.The caller indicated that the surgeon tried to find the piece but at the time of the call had not found it.The intuitive surgical, inc.(isi) technical support engineer (tse) noted that this piece was made of ultem and is not radiopaque.The surgeon was going to determine how to proceed.Isi made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
PERMANENT CAUTERY HOOK
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key13565423
MDR Text Key286518161
Report Number2955842-2022-10306
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874112311
UDI-Public(01)00886874112311(10)N10191125
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470183-14
Device Catalogue Number470183
Device Lot NumberN10191125 0053
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient Age57 YR
Patient SexFemale
Patient Weight138 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-