• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Davol Inc., Sub. C.R. Bard, Inc. -1213643 Kugel Patch; Surgical Mesh

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Davol Inc., Sub. C.R. Bard, Inc. -1213643 Kugel Patch; Surgical Mesh Back to Search Results
Catalog Number 0010105
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); Scar Tissue (2060); Insufficient Information (4580)
Date of Event 09/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol kugel hernia patch on (b)(6) 2008.As reported, the patient is making a claim for an adverse patient outcome against the kugel hernia patch.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient sustained personal injury, experienced emotional distress and the device was defective.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol kugel hernia patch on (b)(6)2008.As reported, the patient is making a claim for an adverse patient outcome against the kugel hernia patch.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient sustained personal injury, experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6)2008 ¿ patient was diagnosed with left indirect inguinal hernia thereby underwent open repair with the implant of kugel patch.Per op notes, ¿the ilioinguinal nerve was dissected free.The hernia sac was noted.A medium oval kugel patch was placed in the preperitoneal space and secured to the inguinal ligament using sutures.¿ (b)(6)2020 ¿ patient was diagnosed with chronic left groin pain thereby underwent open repair with the removal of mesh.Per op notes, ¿the mesh was thick and heavy, found underneath the direct space in preperitoneal position.The mesh was densely stuck to the epigastric vessel and to spermatic cord structure.Part of mesh appeared to be in intensely scarred into peritoneal surface.The mesh was excised.¿.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-mar-2008) is considered to be the best estimate.Updated fields: a2, a4, b3, b4, b5, b7, d4 (product catalog no., corporate lot no., udi no., expiry date), d.6b (date of explant), e1, g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Kugel Patch
Common Device Name
Surgical Mesh
Manufacturer (Section D)
Davol Inc., Sub. C.R. Bard, Inc. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
Bard Shannon Limited 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
+1(401)825-8495
MDR Report Key13566665
Report Number1213643-2022-91073
Device Sequence Number5819444
Product Code FTL
UDI-Device Identifier00801741016370
UDI-Public(01)00801741016370(17)130428(10)HUSC0421
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K963141
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2013
Device Catalogue Number0010105
Device Lot NumberHUSC0421
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer08/05/2026
Initial Report FDA Received Date02/21/2022
Supplement Report FDA Received Date08/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight203 LB
-
-