MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD Back to Search Results
Model Number 7713513
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient implanted with a screw in a previous surgery.It was reported that there was screw loosening at the hcp's 6-week post op appointment.There were no patient symptoms or complications as a result of the event.There was no additional surgery or treatment/ revision surgery/ delay in overall procedure time/ hospitalization necessary as a result of this event.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H6: updated eval.Code result post image review.In the first image of lateral x-ray of c4-7 acdf, kyphotic deformity present.Plate does not contact anterior vertebral column.In the second image, plate appears fixated to anterior spine.No obvious hardware malfunction identified.The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key13567223
Report Number1030489-2022-00164
Device Sequence Number4019172
Product Code KWQ
UDI-Device Identifier00643169329256
UDI-Public00643169329256
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K141632
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7713513
Device Catalogue Number7713513
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date02/21/2022
Supplement Report FDA Received Date06/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
-
-