ZIMMER BIOMET, INC. ECHO POR FMRL NC 9X125MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870)
Date of Event 01/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient underwent a left hip revision approximately 1 month post implantation due to a closed nondisplaced oblique fracture of left femur from a mechanical fall.No additional information.
 
Additional Manufacturer Narrative
(b)(4).Reported event was confirmed due to the review of medical records.Medical records were not provided for the revision that occurred on (b)(6) 2022.Medical notes for the initial procedure state a lateral femoral cortical perforation occurred.Device history record (dhr) was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO POR FMRL NC 9X125MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13568081
Report Number0001825034-2022-00375
Device Sequence Number17658055
Product Code KWA
UDI-Device Identifier00887868260407
UDI-Public(01)00887868260407(17)310304(10)735450
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K070274
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number192009
Device Lot Number735450
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Initial Report FDA Received Date02/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age45 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-