| Model Number |
71436128 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation. a visual inspection of the returned device confirmed the stated failure mode.The journey uni tib tr ins sz 1-2/8mm is chipped rendering device inoperative.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.A supplemental mdr will be sent with the complete results of investigation.Internal complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a journey uni tib tr ins size 1-2/8mm was worn and no longer slides into the tibial tray due to warped edges.It is unknown whether the event happened during surgery and if there was a patient involvement.Upon investigation, it was found that the trial is chipped.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms a piece of the device is missing.The missing piece was not returned.The device has multiple scratches, burrs and deep gouges in the plastic.The device shows signs of significant wear and use.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it cannot longer fit its purpose, the contribution of the device to the reported event could be corroborated.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|