The manufacturer was previously contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged visualization of black particles in filter, shortness of breath and cough.There was no report of serious or permanent patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for investigation.An internal visual inspection of the device was completed by the manufacturer.The manufacturer found evidence of dust contaminant inside the blower box at the location of the filter, on the pca, on the blower box, top enclosure, front panel, rear panel, bottom enclosure, the o-ring of the rear panel, on the blower cap, the blower, blower isolators, and the blower seal.The manufacturer also found evidence of keratin-like substance on the blower box outlet, blower seal, and contaminant on the blower seal.The manufacturer found no evidence of sound abatement foam degradation/breakdown.The manufacturer powered up the device and confirmed airflow.The device's downloaded event log was reviewed by the manufacturer and found no error logged.The manufacturer concludes there was evidence of multiple contaminations from the device.The manufacturer confirmed there was no evidence of sound abatement foam degradation/breakdown.Section d9, g3 and h6 has been updated/corrected in this report.
|