• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN Back to Search Results
Model Number AB9U14150090
Medical Device Problem Codes Device Damaged by Another Device (2915); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use abre self-expanding stent along with non-medtronic 11 fr sheath and 0.035" glidewire during procedure to treat a moderately calcified fibrous lesion in the left distal common femoral vein (cfv).No embolic protection was used.There was no damage noted to packaging.There was no issue noted when removing the device from hoop/tray.The device was prepped per ifu with no issues identified.Stent deformation occurred in vivo post deployment.It was reported that a non-medtronic 16x6x80 pta with manual aspiration.Post dilatation may not have been fully deflated and got caught on end of caudal stent.Balloon did not pop and was safely removed with stent in place.The 14x150 abre stent was deformed by the non-medtronic balloon.Balloon dilated deformed stent and realigned deformed 14x150 with a 16x120 abre stent.16x120 abre stent was implanted to treat the deformed stent.No issues reported post dilation of this stent.The vessel was pre dilated with a 16x5x80 pta balloon.The device passed through a previously deployed stent.There was no resistance encountered when advancing the device.No patient injury was reported.
 
Additional Manufacturer Narrative
Image analysis: three images were returned for evaluation.In the first image you can see the balloon inside the stent and the stent been fully deployed.In the second image you can see a wire inside the stent and the stent appears to be deformed.The third image is unclear.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABRE VENOUS SELF-EXPANDING STENT SYSTEM
Common Device Name
STENT, ILIAC VEIN
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13568709
Report Number2183870-2022-00056
Device Sequence Number5951192
Product Code QAN
UDI-Device Identifier00643169796324
UDI-Public00643169796324
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P200026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/05/2024
Device Model NumberAB9U14150090
Device Catalogue NumberAB9U14150090
Device Lot NumberB230011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date02/21/2022
Supplement Report FDA Received Date05/12/2022
Date Device Manufactured07/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-