• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EIT EMERGING IMPLANT TECHNOLOGIES GMBH TRIAL INSERTER SH CONNECTION; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EIT EMERGING IMPLANT TECHNOLOGIES GMBH TRIAL INSERTER SH CONNECTION; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number TFT20101
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in germany as follows: it was reported that on (b)(6) 2021, the product was found to be defective.This report is for one (1) implant inserter sh connection.This is report 2 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: a device history record (dhr) review was conducted: visual inspection: the complaint device trial inserter sh connection was returned to cq west chester for investigation along with implant inserter pin and knob.The trial shaft had the implant inserter knob(tft:30101b) locked on to its end and the opening of the shaft for the pin appeared deformed.Functional test: the implant inserter knob could not be removed from the trial shaft as a broken piece of inserter pin was jammed inside the shaft & knob.Can the complaint be replicated - yes, the complaint condition can be replicated.Document /specification review: based on the date of manufacture, current and manufactured revision of drawings were reviewed.Dimensional inspection: complaint relevant dimension cannot be measured without the destruction of the device.Complaint confirmed: complaint can be confirmed based on the available information.Conclusion: the implant inserter pin and knob were used with trial inserter shaft.The implant inserter might have been incompatible with the trial shaft leading to the complaint issue.This is could be a user error.But a definitive root cause was not determined during investigation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities device history lot - part # tft20101 synthes lot # e19di1432 supplier lot # e19di1432 release to warehouse date: 02 jun 2020 supplier: (b)(4) no ncr's were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAL INSERTER SH CONNECTION
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
eisenbahnstrasse 84
wurmlingen D-785 73
DE  D-78573
Manufacturer (Section G)
BRIDGEWATER DISTRIBUTION
50 scotland boulevard
bridgewater MA 02324
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13568935
Report Number3013730328-2022-00014
Device Sequence Number6062148
Product Code MAX
UDI-Device Identifier04260557775578
UDI-Public(01)04260557775578
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K170503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFT20101
Device Catalogue NumberTFT20101
Device Lot NumberE19DI1432
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Initial Report FDA Received Date02/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
IMPLANT INSERTER SH CONNECTION
Patient SexUnknown
-
-