This is filed to report leak, air embolism, delay and prolonged hospitalization, and medical intervention.
It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3-4.
A steerable guide catheter (sgc) was advanced to the mitral valve; however after the dilator was retracted from the sgc, air was observed in the guide.
Additionally, the echocardiographer observed air in the apex of the heart causing a clinically significant delay in the procedure.
The air was then immediately aspirated from the guide and a pigtail catheter was advanced to the heart to fully remove the air from the heart.
Then the procedure continued and the clip delivery system (cds) was inserted into the sgc.
However, prior to the clip reaching the heart, the patient's blood pressure quickly dropped and the patient was resuscitated.
The cds was immediately removed with the clip attached and the sgc was retracted from the heart into the vena cava inferior.
The procedure was paused to place an impella pump and perform cardial reanimation.
When the impella was positioned and taking over the pumping, a decision was made to continue with the mitraclip procedure.
A new sgc was being prepared for use due to the origin of the air seemed unclear.
During preparation of the new sgc, the patient's blood pressure dropped again and the patient required cardial reanimation again.
The initial sgc was removed from the patient and the mitraclip procedure was aborted.
The echocardiographer assumed that some air went into the coronaries as the right heart did not have any ejection fraction (ef) anymore, an extracorporeal membrane oxygenation (ecmo) was installed to keep the patient stable.
The patient is currently stable, but in critical condition in the intensive care unit and will remain hospitalized.
No clips were implanted, and mr is 3-4.
No additional information was provided.
|